protection, PWM pin voltage fault indication terminal outputs high level.
3.19.5
Current monitoring (I Monitor)
The 0-10V analog quantity output signal of current monitoring output terminal
represents input current to which the terminal belongs from 0 to full range. An
external voltmeter or oscilloscope can be connected to display input current
change.
3.20 Auto test function
The IT8912E electronic load delivers strong auto test functions, which can
analog several tests. A total of 10 groups of test files can be edited, and each
group test file has 10 steps. Therefore, a maximum of 100 steps can be edited
and saved in EEPROM.
NOTE
In the following editing procedures, "Y indicates selected status. To cancel selected status,
press numeric key of corresponding step again.
Operation steps
1. Press
[Shift]
+ 6 keys.
PROGRAM
Run Recall Edit
2. Operate
key and move to Edit. Press
[Enter]
key to enter editing
test files
EDIT PROGRAM
Active Sequence=0987654321
3. Press numeric key to select test step and press
[Enter]
key. Active
Sequence= 09876543YY (indicating that 1/2 step has been selected).
EDIT PROGRAM
Active Sequence=09876543YY
4. Select whether pause is necessary for these two steps. If step 2 is to be
paused, press 2 key. If not, directly press
[Enter]
key.
EDIT PROGRAM
Pause Sequence=
□□□□□□□□Y1
5. Select whether short-circuit test is necessary for these two steps. If step 1
is to be tested, press 1 key. If not, directly press
[Enter]
key.
EDIT PROGRAM
Short Sequence=
□□□□□□□□2Y
6. Set loading time of step 1. If 2S is required, directly press 2 key on the
board. Press
[Enter]
key.
EDIT PROGRAM
SEQ01 On Time=2S
7. Set unloading time of step 1. If 2S is required, directly press 2 key. Press
[Enter]
key.
EDIT PROGRAM
SEQ01 Off Time=2S
8. Set test delay time of step 1. If 1S is required, directly press 1 key. Press
[Enter]
key. Tpf is delay time before measurement.
EDIT PROGRAM
SEQ01 P/F Delay Time=1S