Function
and Features
Copyright ©ITECH Electronic Co., Ltd.
14
NOTE
When the set current is within small range, current display resolution will be improved.
3.2.2
Constant voltage operation mode (CV)
Under CV mode of IT8900, you can set limit current (Limit). When the set limit
current is higher than the input current, the electronic load will work under
constant voltage; when the set limit current is lower than the input current, the
electronic current will be loaded with set limit current and the voltage is input
voltage. As shown in Fig. 3-2.
Fig. 3-2 Voltage-Current Relation Schema under CV Mode
Under CV mode, the electronic load provides three ways to modify constant
voltage.
⚫
Rotate adjusting knob to set constant voltage value.
⚫
Use numeric keys to input voltage value and press
[Enter]
key to confirm
set constant voltage value.
⚫
Use
moving cursor and press
to adjust values at
corresponding positions.
Operation steps
Set voltage range.
1. Press
[CV]
key and
[Shift]
+
[CV]
to enter parameter setting screen.
Constant Voltage
Range=500.00V
2. Set maximum working voltage value and press
[Enter]
for confirmation.
Constant Voltage
Range=500.00V
3. Set maximum current value and press
[Enter]
key.
Constant Voltage
▲
High=15.000A
▼
4. Set minimum current value and press
[Enter]
key.
Constant Voltage
▲
Low=0.0000A
▼
5. Set high and low rate and press
[Enter]
key.
Constant Voltage
▲
High-Rate Low-Rate
6. Complete parameter setting.
Load current
Load
Voltage
I
V
Set voltage
CV mode