Function
and Features
Copyright ©ITECH Electronic Co., Ltd.
24
11. Enter the dynamic test mode.
10.0000V 0.0000A
0.00W 0 TRAN
12. Press
[On/Off]
key to open input and press
[Shift]
+
(Triggering key)
The load will continuously switch between A and B value. Time of
operations is shown at bottom right.
13. Press
[CC]
/
[CV]
/
[CR]
/
[CW]
key or any composite function key to exit
dynamic test function. Repeat Steps 1-12 to continue parameter setting
and operation of dynamic test.
3.11.2
Pulse mode
Under pulse mode, after enabling dynamic test operation, the load will switch to
B value after receipt of a trigger signal and switch back to A value after
maintaining B for pulse width time.
Taking CC mode as example (operations under other modes are similar), when
output voltage and current of the tested instrument are 10 V and 3 A
respectively, the load current will switch from 1 A and 2 A. Set the dynamic test
parameters and perform the test as follows:
1. Press
[Shift]
+ 2 keys.
TRANSITION
On Off
2. Operate
key and move to On. Press
[Enter]
key and select
Pulse. Press
[Enter]
key (The Trig lamp that indicates VFD screen status
is on)
TRANSITION
Continuous Pulse Toggle
3. Operate
key and select high range and low range. Move to the
High-Rate and press
[Enter]
key.
TRANSITION
High-Rate Low-Rate
4. Set ascending slope and press
[Enter]
key.
TRANSITION
Rise up=30.000A/uS
5. Set descending slope and press
[Enter]
key.
TRANSITION
Rise down=30.000A/uS
6. Set A value and press
[Enter]
key.
TRANSITION
Level A=1A
7. Set B value and press
[Enter]
key.
TRANSITION
Level B=2A
10A
5A
Pulsed Transient Operation
10ms
TWD
10ms
TWD
TRIG
TRIG