4.20
CAP-CAL reference capacitor (code PII40175)
The following image exhibits the CAP-CAL reference capacitor:
Figure 33 - CAP-CAL reference capacitor
Purpose of the reference capacitor is to check the actual calibration of TD 5000 for the measures of the capacitance and
Tan(δ).
The CAP-CAL includes an extremely high accuracy high voltage and low Tan(δ) capacitor. The device includes also 4
resistances to be connected to the capacitor in series mode, to generate some known Tan(δ) values.
The following image exhibits the connection of the CAP-CAL to the TD 5000:
Figure 34 - Connection of the CAP-CAL to the TD 5000
A calibration certificate is issued by ISA laboratory and it is supplied together the CAP-CAL.