Invasix BodyTite™ Service Manual DO600463F
0344
Page 33 of 52
#
Failure
Description
Action
fault
12
RF Energy card
fault
Recalibrate RF card
Replace RF card
13
Always Bad
Coupling
Replace Hand piece
Recalibrate RF card
Replace RF card
Check wiring from RF card to Hand piece connector
14
Key does not
respond
Replace LCD adaptor card
15
Footswitch does
not respond
Replace footswitch
Check controller to footswitch wiring
Replace Controller card
16
Hand piece LCD
is blank
Replace hand piece
17
2nd RF is not
produced
Replace hand piece
17
No Vacuum
Replace hand piece
Содержание BodyTite