System Specifications
Input Power
Main Line Frequency (nominal)
Input Voltage (nominal)
Input Current (rms)
50-60 Hz
100-240 VAC
1.5A,
Operating Parameters
Ambient Temperature Range
Relative Humidity
Atmospheric Pressure
Warm-up Time
15° – 35° C [59° – 95° F]
30% to 80%, non-condensing
700 kPa to 1060 kPa
If transported or stored at temperatures outside the operating
temperature range, allow one hour for the to reach room
temperature before use.
Transport and Storage
Ambient Temperature Range
Relative Humidity
Atmospheric Pressure
-20° – 65° C [35° – 131° F]
0% to 80%, non-condensing
500 hPa to 1060 hPa
Dimensions
System
Handpiece cable
36cm W x 36cm D x 100cm H
250 cm
[14.2’’ W x 14.2’’ D x 40’’ H]
[100``]
Weight
System
Applicator
15 Kg
1 Kg
[33 lbs.]
[2.2 lbs.]
Output Parameters
Output Configuration
1
st
RF
Maximum Output Power
Output Frequency
V
p-p
max
Crest Factor * (Rated Load)
2
nd
RF
RF V
p-p
max
RF Maximal sequence duration
Output Frequency
Isolated Output
75 [W] @ 300[Ω]
1[MHz] ± 2%
600[V]
2 ± 2%
2000V
5μsec
1.7[MHz] ± 2%