Invasix BodyTite™ Service Manual DO600463F
0344
Page 31 of 52
H001 Fan Fault
:
The fan status signal indicates a fault.
Reason
: The fan does not rotate.
H002 Hand piece not connected
:
The application software fails to detect a hand-piece.
Reason
: The hand piece is not connected (most probable) or hand piece
controller is not functioning.
H003 RF Card Fault
:
RF card does not behave as expected,
Reason
: The RF monitoring system shows unexpected values while no RF is
delivered.
H004 Used hand piece indicated:
Software detected a used Hand Piece
Reason
: The BodyTite HP has made more than 150KJ or FaceTite HP has
made more than 50KJ.
H005 System Memory Fault:
The software failed to read system parameters
Reason
: The system parameters (RF calibration) are corrupted. It may appear
in a system that was never calibrated, corruption after calibration was never
faced.
H006 Low Temperature:
Message is displayed on screen if the temperature does not rise by ½ ºC
during 10 seconds. RF delivery is stooped, but it does not cause system shut-
down.
Reason
: The temperature sensor is probably unconnected. No error number
because it is not a shut down error
H007 Temperature Sensor Fault:
The measured temperature is over 60 ºC (probably a short in the thermistor)
Reason
: The temperature sensor is probably shorted.