Top Assy Disassembling
Remove front cover as shown in the Front cover removal section
Remove the Keypad flat cable from the controller card
Remove the LCD flat cable from the controller and release it from the ferrite
Open the 4 screws attaching the Top Assy to the Top Panel
Open the 2 screws attaching the Top Panel to the chassis located on the front side
opening top part
Tilt forward the Top Assy
Remove the two flat cables from the LCD adaptor
Remove the connector bracket from the Top Assy by removing the 2 brackets holding
it in place
Pull up the Top Assy and let the two flat cables go through their opening
Top Assy Assembling
Position the top Assy to the top panel and tilt it so the connector bracket can be
attached to it using the two holding brackets, use four screws to attach it
Connect the two flat cables to the adaptor card and push the free edge through the
openings
Lay down the top Assy on the top panel
Connect the keypad flat cable to the controller card
Rout the LCD flat cable through the ferrite and connect it to the controller card
Attach the top Assy to the top panel using the four screws