Invasix BodyTite™ Service Manual DO600463F
0344
Page 32 of 52
Section 8 - Troubleshooting
#
Failure
Description
Action
1
No Display (No
Backlight, No
fan)
Make sure power cable is all the way in
Check fuses
Check AC filter card
Check Mains wiring (Including switch operation)
Check Power supply output
2
No Display (No
Backlight, Fan
OK)
Check DC power connection to controller
Replace LCD Flat cable
Replace LCD Flex cable
Replace LCD adaptor card
Replace controller card
3
Display no
graphics and
characters
(Backlight is On)
Replace controller card
4
No Software
plug
Replace Software plug
Replace Controller card
5
SW Memory
Fault Check
Sum
Verify that the memory plug is tight. Turn the System on.
If massage reappears, replace memory plug..
6
Fan Fault
(Display is OK)
Verify that the fan is not blocked by an external object. If the
fan is blocked, remove the object.
Check DC power connection to controller
Check fan wiring
Replace Fan
7
Hand Piece
Connection
Fault
There is no hand piece connected
Check wiring from controller to Hand Piece
Bad Hand Piece
8
Hand Piece
Memory Fault
Replace Hand Piece
9
Temperature
sensor fault
(only for
BodyTite)
Replace Hand Piece
Check wiring from RF card to Hand Piece connector
Replace RF card
11
System memory
Recalibrate RF card
Содержание BodyTite