Invacare® Bora® / Spectra® XTR Series
Problem
Other symptoms
Possible cause
Solution
Documentation
Remote defective
Replace remote
See corresponding
electronics manual
Mobility device runs
too slowly
None
Batteries defective
Replace batteries
See
5.3 Tightening Torques
CAUTION!
Risk of damage to mobility device due to
improperly tightened screws, nuts or plastic
connections.
– Always tighten screws, nuts etc. to the stated
tightening torque.
– Only tighten screws or nuts which are not
listed here fingertight.
The tightening torques stated in the following list are
based on the thread diameter for the nuts and bolts for
which no specific values have been determined. All values
assume dry and de-greased threads.
Thread
Tightening Torque in
Nm ±10 %
M4
3 Nm
M5
6 Nm
M6
10 Nm
M8
25 Nm
M10
49 Nm
M12
80 Nm
M14
120 Nm
M16
180 Nm
5.4 Service Plan (Once a Year)
CAUTION!
Risk of injury and damage to property, if the maximum speed reduction on a mobility device with a lifter does
not function correctly
The mobility device’s control unit must reduce the maximum possible speed as soon as the lifter is raised.
– Test the maximum speed reduction for correct function after any maintenance work or modifications
to the mobility device.
Component
Check
Remedy
Notes
Risk of damage to
armrests
Replace shrouds, if
damaged
Armrests
Armrest fixings
Tighten screws
Risk of damage to
clothes-guards
Replace clothes-guards, if
damaged
Clothes-guard
Clothes-guard fixings
Tighten screws
Seat lock
Seat lock defective
Replace seat lock
Tilt
Check SL retainer clip for
correct fit
Replace SL retainer clip, if
necessary
Risk of damage to backrest
Seams
Fixing
Check cable
Power recline (if fitted)
Check function
Replace parts, if damaged
Tighten screws
Replace actuator cable, if
necessary
Tighten screws
Frames (chassis) / battery
mounting
Check fixings, welded
seams and battery
mounting
Replace components, if
necessary
14
1528338-Q