5 Service
5.1 General Warning Information on
Installation Work
CAUTION!
Risk of damage to vehicle
Collisions can be caused if the adjusting washers
are removed during fitting work to the drive
wheels. Adjusting washers are often fitted
between the drive shaft and the wheel hub to
even out tolerances. If these adjusting washers
are removed and not replaced again, collisions
can be caused.
– Always replace the adjusting washers exactly
as they were before you started dismantling.
CAUTION!
Risk of injury and damage to property, if
the maximum speed reduction on a mobility
device with a lifter does not function correctly
The mobility device’s control unit must reduce
the maximum possible speed as soon as the
lifter is raised.
– Test the maximum speed reduction for
correct function after any maintenance work
or modifications to the mobility device.
5.2 Troubleshooting
5.2.1 Operational Faults
Proceed as follows if you have any problems:
1.
First assess the possible cause of the problem using
the following table.
2.
Check the remote status display. Evaluate the flash
error code.
3.
Carry out the necessary checks and repairs as
recommended in the following table.
The various power modules can be fitted in connection
with different remotes in the mobility device. Rectification
of operational faults depends on the power module
fitted. The power modules used are described in the
corresponding controls manual.
The tables for rectification of operational faults
listed in the following chapters are only an excerpt
from the original manufacturer's manuals. You can
obtain the original manuals from Invacare.
5.2.2 Drive Fault Diagnosis
Problem
Other symptoms
Possible cause
Solution
Documentation
The remote status
display illuminates
normally and shows
an error code.
Drive motors
disengaged
Engage drive motors
See corresponding
remote manual
Batteries defective
Replace batteries
See
Completely discharge
battery
Pre-charge batteries
See user manual
Power supply to
remote interrupted
Check main fuse
See
Remote status display
does not illuminate
Remote defective
Check cables between
the modules for
loose connections and
damage
See
Mobility device will
not start
Remote status display
blinking
Various causes
Assess error code
See corresponding
remote manual
Batteries defective
(unstable voltage)
Replace batteries
See
Replace motor(s)
See
Mobility device
judders in drive mode
None
Drive motor(s)
defective
Replace carbon
brushes
None
Batteries defective
Replace batteries
See
Batteries not being
charged
LEDs blinking on
charging unit
Charging unit defective
Replace charging unit
See user manual
1528338-Q
13