Invacare® Pronto™ M41 Series
1.
Remove two mounting screws (2) that secure posture
belt to seat frame.
2.
Remove two halves of the posture belt (3) from rear
seat frame.
3.
Reposition two new posture belt (3) halves underneath
seat rails.
4.
Install two mounting screws (2) to secure posture belts
(3) to seat frame. Securely tighten.
Fig. 7-8
7.8 Replacing Footboard Assembly
WARNING!
Pinch point may occur when rotating the footboard assembly.
Removing Footboard Assembly
1.
Remove quick release pin (2) that secures footboard
assembly (4) to wheelchair frame by depressing button
while sliding pin out.
2.
Remove footboard assembly (4) from wheelchair frame.
Fig. 7-9
Installing Footboard Assembly
Make sure the detent balls (3) of the quick-release pin (2)
are fully released beyond the outer edge of the tube (5)
before operating the wheelchair. Otherwise, injury and/or
damage may result. Keep detent balls clean.
WARNING!
Make sure the detent balls (3) of the
quick-release pin (2) are fully released beyond
the outer edge of the tube (5) before operating
the wheelchair. Otherwise, injury and/or
damage may result. Keep detent balls clean.
1.
Position the footboard assembly (4) onto wheelchair
frame so that mounting hole (1) in wheelchair frame
aligns with desired mounting hole (1) in footboard
assembly (4).
2.
Install quick release pin (2) by depressing button while
sliding pin in. Ensure that detent balls (3) are engaged
with outer edge of tube (5).
48
1568396-C