Invacare® Pronto™ M41 Series
Component
Check
Remedy
Notes
Power legrests (if
fitted)
Check cable
Check contacts
Check functions
Replace cable if
necessary.
Battery mounting
Check battery support
and mounting belts
for damage
Replace if necessary.
Check batteries for
damage
Replace batteries if
necessary.
See
and
Damaged Batteries Correctly, page
40
Check battery voltage
Charge batteries.
See user manual.
Batteries
Check contacts and
terminals
Clean contacts and
terminals.
See
Remote, status LED
flashing
Evaluate error/flash
code.
See
.
Fixings
Tighten fixings,
replace if necessary.
Cables and connecting
plugs
Replace damaged
cables and tighten
connecting plugs, if
necessary.
Replace joystick knob
if necessary.
Joystick function
Replace remote if
necessary.
Remote / power
module
Power supply
Replace damaged
cables and tighten
connecting plugs, if
necessary.
Chair configuration
Check chair
configuration version
Update software
if newer version
available.
See LiNX service manual.
Screws (every 6
months)
Check screws for tight
fit
Tighten screws if
necessary.
6.6 Overview
Under Rear Shroud/Seat
Depending on seat variant
(1) Power module
(2) Actuator module (optional)
Fig. 6-1
24
1568396-C