Guidance and manufacturer’s declaration - electromagnetic immunity
The Integra Padgett Electric Dermatome Models B and PI are intended for use in the
electromagnetic environment specified below. The customer or the user of the Integra
Padgett Dermatome Models B and PI should assure that they are being used in such an
environment
Immunity Test
IEC 60601 test level
Compliance
level
Electromagnetic Environment -
Guidance
Electrostatic
discharge, (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV
±8 kV
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/
output lines
±2 kV
±1 kV
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
±1 kV
±2 kV
Mains power quality should be
that of a typical commercial or
hospital environment
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical commercial or hospital
environment.
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines IEC
61000-4-11
<5 % UT
for 0,5 cycle
<40 % UT
for 5 cycles
<70 % UT
for 25 cycles
<5 % UT
for 5 sec
><5 % UT
for 0,5 cycle
<40 % UT
for 5 cycles
<70 % UT
for 25 cycles
<5 % UT
for 5 sec
Mains power quality should
be that of a typical commercial
or hospital environment. If the
user of the Integra Padgett
Electric Dermatome Models
B and PI requires continued
operation during power mains
interruptions, it is recommended
that the Integra Padgett Electric
Dermatome Models B and PI be
powered from an uniterruptible
power supply or a battery.
NOTE UT is the a. c. mains voltage prior to application of the test level.
Содержание Padgett B
Страница 30: ...29 This page is intentionally left blank...
Страница 31: ...30 Mod les de dermatome lectrique B et PI Integra Padgett GUIDE DE L UTILISATEUR FR Fran ais...
Страница 60: ...59 This page is intentionally left blank...
Страница 61: ...60 Integra Padgett Dermatomo elettrico modello B e PI GUIDA DELL UTENTE IT Italiano...
Страница 90: ...89 This page is intentionally left blank...
Страница 91: ...90 Integra Padgett Elektrisches Dermatom Modelle B und PI BENUTZER HANDBUCH DE Deutsch...
Страница 120: ...119 This page is intentionally left blank...
Страница 121: ...120 Dermatoma el ctrico Integra Padgett modelos B y PI GU A DEL USUARIO ES Espa ol...
Страница 150: ...149 This page is intentionally left blank...
Страница 151: ...150 Integra Padgett elektrische dermatoom modellen B en PI GEBRUIKERS HANDLEIDING NL Nederlands...