User manual |
AED mode
89
AUDITIVE INFORMATION SIGNAL (BEEP)
DESCRIPTION
Metronome for CPR compressions
The equipment beeps at a frequency of 100 beeps
per minute, helping the user in performing the cardiac
compressions.
The equipment also has verbal information signals. The differentiation of verbal
information signals and verbal alarm signal is given in two ways:
Sound power
Verbal information signals will always be at least 6 dB(A) below the set to the alarm
signal.
Messages content
Verbal information signals have the following messages:
• Analyzing heart rhythm.
• Perform CPR for two minutes.
• Treatment recommended – do not touch the patient.
• Shock is not recommended.
• Put the paddles on the patient’s chest.
• Energy is being discharged internally.
• Don’t touch the patient.
• Children's paddles.
• Treatment successfully.
• Give two breaths.
• Check the patient reparation or pulse if you cannot detect either, perform CPR
for two minutes.