Battery use
ATTENTION: observe the battery charge maintenance instructions.
First use
Before using the DualMax for the first time, the equipment must receive a full
battery charge. In order to do this, the equipment needs to be connected to an
electric current for at least eught hours.
Occasional use
Even when disconnected (stand-by), the DualMax executes internal routines
checking the status of the equipment. In spite of this procedure entailing a low
power consumption, the battery charge may be consumed.
Therefore, whenever the device has not been connected to an electric current for
more than 20 days, it is advisable to execute a full battery charge. If this procedure
is not performed, there is a risk of draining the battery and consequently being
unable to use the DualMax in its portable configuration (not connected to the
electric current).
Storage
The battery must be removed from the equipment in case it is stored or not used.
Replacement
Every battery has a determined lifetime, which is the possible quantity of full charge
and discharge cycles, without loss of performance. When the appliance has a drop
in performance of the battery, with low autonomy, request a new unit from Instramed
technical assistance.
The battery can be replaced following the procedures described on the chapter
“Care and maintenance”.
Battery replacement is recommended every 02 (two) years or when the runtime is less
than 01 (one) hour.
ATTENTION: the equipment is capable of operating with only one battery
connected.
ATTENTION: when servicing the batteries, it is obligatory to change only
one battery at a time, leaving the other battery in the device, so that it can
continue to operate at that moment.