User manual |
RTC - Real Time Check
167
It is possible to interrupt the test with a click on the e-Jog button.
After completing the test, the equipment will send the collected data to the RTC
Control Center. At this time, it is not possible to turn off the device. If the connection
with the RTC Control Center cannot be established, the device will cancel sending
the data after 10 seconds. After this procedure, the device will save the collected
data and turn off automatically.
NOTE: the test of the defibrillator module will not be carried out if:
•
The equipment is connected to the computer by USB cable.
•
The paddles are not on the support or are disconnected from the device
(external adult paddles).
•
The battery has a charge of less than 10% and the equipment is not
connected to a power outlet.
Manual test
Follow the instructions on screen to start the test. Or click "cancel" to cancel the
test.
Once started, the manual test goes through the following steps:
• Test of the connection with the ECG board.
• Test of the defibrillator module.
• Check of the battery level.
• Check of the connection with the power outlet.
It is not possible to interrupt the manual test. After completing the test, the
equipment will send the collected data to the RTC Control Center. At this time, it is
not possible to turn off the device. If the connection with the RTC Control Center
cannot be established, the device will cancel sending the data after 10 seconds.
After this procedure, the device will save the collected data and return to the mode
defined by the rotating switch on the DualMax.
Error BEEPS
When there is an error in the last RTC test, the equipment emits three beeps
in sequence, which can be identified by the user from a distance. This beeping
sequence will be repeated every three minutes until the failure is solved.