User manual |
Monitor mode - ECG
105
ECG Setup - Principal
1
2
4
3
5
9
10
8
7
6
1 - ECG response
Select the ECG numeric update response. You can select from “SLOW, “NORMAL”
and “FAST” responses.
NORMAL:
used for most patients, this mode uses 16 beats to define the average.
FAST:
used when the operator needs faster responses. It is affected by the patient's
movements, using 8 beats to define the average.
SLOW:
less affected by the patient's movements. However, you must pay attention
to the slow response of the heart rate variation. This mode uses 32 beats to define
the average.
2 - ECG cable
It allows selecting the ECG monitoring by cable with 3 lead, 5-lead or 10-lead.
Using the 3-wire ECG cable in the SDP (Sudden Death Prevention) makes
it functional only in the DII lead.