User manual |
Specifications and safety
203
Environmental specifications
Temperature:
Operational: 0 to 50ºC.
Storage: - 20 to 50ºC.
Humidity:
Operational: 10 to 95% RH, without condensation.
Storage: 10 to 95% RH, without condensation.
Altitude:
Recommended to use in altitudes lower than 2000 meters.
WARNING: if the DualMax is used outside these conditions, 15 through 30
minutes will be required to stabilize the system so that functioning failures
do not occur.
Defibrillator
Waveform:
Biphasic truncated exponential. Wave shaped parameters
adjusted according to the patient's impedance.
Shock application:
By means of multifunction pads or external adult/child pads.
Scales for adult/external
defibrillation:
1, 2, 3, 4, 5, 6, 7, 8, 9, 10, 20, 30,50, 80, 100, 150, 200, 250,
300 and 360 J.
Energy option delivered limited to 200 J.
Maximum energy limited to 50 J with child pads.
Charge auto-sequencing:
When enabled, it charges power previously set by the
user for the first, second and third shocks, with no need to
manually adjust the selector.
Energy selection:
Selector switch in front panel.
**Optional: energy selection in the external pads (consult
availability)
.
Charge command:
Button in front panel, buttons in external pads.
Shock command:
Button in front panel, buttons in external pads.