Defibrillator mode
77
3 - Third shock
Sets energy levels for the third shock of the Auto Seq mode. The user can select
energy levels from 10 to 360 J. Preset values are 150 J for the first shock and 200 J
for the next shocks.
In case child pads are used, the equipment will automatically limit the charge to
50 J. When there is use of child pads followed by the use of adult pads, the last
configured values for adult pads will be recovered.
The charge and shock delivery processes are operated manually and after every
delivery, the energy levels are updated.
4 - Back/Exit
“BACK” to configuration menu or “EXIT” to go to the monitoring screen.