Event and data storage
18
Data storage
The DualMax creates an event list for each patient observing the following criteria:
Automatically
- If DualMax stays turned off for 30 seconds more.
Manually
- Via the events setup menu.
The equipment stores data from up to 100 patients and 800 generated events. In
each event it will be stored the ECG curve 12 seconds before and 12 seconds after
the event generation.
When in AED mode, the equipment stores in its memory the last two ECG hours.
Events stored
The DualMax stores the date, time, and ECG curve of the following situations:
• Defibrillation/cardioversion.
• Stimulation - Pacemaker On/Off.
• "Pause audio" key activation.
• Shock failure.
• Internal discharge.
• Manual event marking.
• Indicated shock or asystole (for equipment with SDP).
• Values out of the the alarm limits.
• Change of pads.
• The events continue to be stored, even when SOUND ALARM of a particular
parameter is turned off. The shutdown time is not registered.
• In case of total loss of energy, the content of the records do not suffer
modification.