User manual |
Care and maintenance
190
Repairs
If it is necessary to send the DualMax for repairs or to clarify other doubts, contact
Instramed at +55 (51) 30738200 or by the e-mails: [email protected].
br and suporte@instramed,com,br. Please, be prepared to inform the equipment’s
serial number.
If possible, use the original equipment’s packaging. If this is not possible, use an
equivalent box that provides adequate protection for the monitor.
Precautions, restrictions and warnings
The DualMax is a device built according to NBR and IEC standards and therefore
offers total safety for patient and operator, However, all safety precautions
described below must be followed.
The monitor's operation can be affected by the presence of electromagnetic
power sources, such as electrosurgical equipment and computer tomography
(CT).
1 - ECG
1 -
To guarantee protection against the effects of a defibrillation, use only the patient
cable that accompanies the equipment.
2 -
If the monitor is used simultaneously with an electric scalpel, position the ECG
electrodes as far as possible from the RF current route, between the surgical
field and the neutral card, Do not use needle type ECG electrode during surgical
procedures.
2 - SpO
2
1 -
The operation of this device may be affected by the presence of electromagnetic
energy sources, such as electrosurgical or CT equipment. It also may be damaged
by the presence of strong ambient light. If necessary, protect the area of the sensor
with a surgical towel.
2 -
Any dyes injected into the blood stream, such as methylene blue, indocaine
green, indigo carmine and fluorescein, may affect the SpO
2
reading precision.
The presence of dyshemoglobin, such as carboxyhemoglobin (in consequence of
carbon monoxide poisoning) or methemoglobin (in consequence of sulfonamide’s
treatment) may affect the SpO
2
reading precision.