96-06434-00-01 revA
Service Manual, Inogen One G3HF
-62-
2.
Remove the Exhaust tube from the Tube Bracket and Muffler. Then unclip the Column
tube from the Inlet Union Barb
Figure 7: Exhaust Tube and Column Tube
3. Locate the two columns tubes attached to the Product Manifold. Release the Column B
Tube from its quick connect by pushing down on the retaining ring while pulling on the
tubing. (Figure 8).
Figure 8: Exhaust Tube and Column Tube
4. Remove the Column A tube from the Product Manifold by pushing down on the
retaining ring of the quick connect while pulling the tubing down (Figure 9). Then
release the accumulator tubing from its quick connect in the same manner.