96-06434-00-01 revA
Service Manual, Inogen One G3HF
-20-
Figure 9: Housing feet and corresponding chassis grooves
7. Place the device on its front side and repeat step 6 to secure the back housing into the
chassis.
8. Replace the two screws that connect the back housing to the chassis and the two screws
that connect the front housing to the chassis. Tighten screws to 3 in-lbs.
9. Return the concentrator to the upright position.
10. Slide the battery back into the chassis. Ensure the battery clip is fully seated.