96-06434-00-01 revA
Service Manual, Inogen One G3HF
-30-
2.
Detach the Inlet tube from the inlet union barb (Figure 6 and 7) and remove the Inlet
tube from the device. Unsnap the Column tube from union barb if necessary.
Figure 6 and 7: Inlet tube and inlet union
Instructions for Replacing the Inlet T Tube:
1. Reinstall the new Inlet tube in the orientation below with the top end of the tube
attached to the manifold, ensuring that the alignment mark is on top of the manifold
barb (See Figure 8). Be sure to feed the inlet tube back under the Accumulator tube.
Seat the tubing as far as possible on to the manifold barb.
Figure 8: Orientation of Inlet T Tube
2.
Attach the bottom end of the new Inlet tube to the inlet union, with the other alignment
mark pointing down on the inlet union barb (Figure 6 and 7). Make sure the tubing is
fully seated onto the inlet union.
3.
Re-install the motherboard.