Instructions for Replacing the Housing:
1. Place the device on its front side and seat the feet of the back housing into the
corresponding grooves on the chassis (Figure 8) then push down and toward the device
to secure the housing in the chassis.
Figure 8: Housing feet and corresponding chassis grooves
2. Replace the two screws that connect the back housing to the chassis (Figure 2). Tighten
screws to 3 in-lbs.
3. Return the concentrator to the upright position
4. Slide the battery back into the chassis. Ensure the battery clip is fully seated.