96-06434-00-01 revA
Service Manual, Inogen One G3HF
-25-
Figure 9: Molded Column Tubes on Feed Waste Manifold
Instructions for Replacing the Column Receptacles:
1. Install the tube back onto the Feed Waste Manifold barb with the clocking line directly on
top of the barb (Figure 9).
2. Reinstall either the Accumulator Sensor Tube routing it around the long molded column
tube and inserting it into the quick-connect fitting on the Accumulator (Figure 8), or the
manifold harnesses by reconnecting the Product Manifold Harness, then the Feed Waste
Harnesses to the motherboard (Figures 6 and 7).
3. Install the column receptacle into the molded column tube, ensuring that the clocking lines
on the tube are directly on top of the receptacle barb (Figure 10). Place the receptacle on
top of the column.
Figure 10: Column Receptacle and Tubing
Short Molded Column Tube
Long Molded Column Tube