96-06434-00-01 revA
Service Manual, Inogen One G3HF
-111-
4. Disconnect the singular end of the feed tube from the inlet union using a small
flathead screw driver to push the tube off of the connector (Figure 4).
Figure 4: Remove feed tube from inlet union
5. Unscrew the two screws on the front of the device and the bottom center screw on the
back of the device that connect the filter body to the chassis (Figures 5 and 6).
Figure 5: Location of front screws
Figure 6: Location of rear screw
Feed tube and inlet