![Inogen One G3HF Скачать руководство пользователя страница 116](http://html1.mh-extra.com/html/inogen/one-g3hf/one-g3hf_service-manual_2064301116.webp)
96-06434-00-01 revA
Service Manual, Inogen One G3HF
-116-
Figure 2: Location of screws
3. With the concentrator still on its front side, slide the back housing down and away from
the device then lift up to remove (Figure 3).
Figure 3: Direction to slide housing
Instructions for Removing the Column Latch:
1. With the device still on its front side, locate the spring nest, spring, and column latch
assembly. Gently pull the spring nest away from the chassis and remove the spring nest
and spring together (Figure 4).