Doc ID: 900434-001
March 2018
9
FaILUres
In case you observe any failures or receive any failure message by your workstation
software without clear remedy instruction please contact the support of your workstation
provider.
9.1.
Drawer failure
9.1.1.
error 10
• check whether the labware matches the labware criteria specified in chapter xxx
•
check whether the opening for the electric eye of drawer detection is open
Fig.27:
opening for electric eye
•
check whether the electric eye of drawer detection is clean
Fig.28:
opening for electric eye
•
reboot the incubator