4.7. 

Labware Use and Detection

The heat transfer is optimized for microplates with  flat bottom to ensure uniform heat 

exchange. If you observe a non uniform heating of other plate types (e.g. round button) 

a special adapter might be needed. INHECO offers to design and manufacture 

customized adapter.
The devices can measure the temperature at the top of the heat plate only. The 

temperature of the liquid in the microplate differs from this. Therefore a heater offset 

must be defined and set before the temperatures in the liquid samples are correct (see 

also command SHO in the firmware command set). 

The heat plate has three integrated sensors for measuring the temperature. The 

temperature of the sensors is averaged. 

Only for Incubator with shaking fuction:

When a labware is placed on the drawer and the door is being closed, the labware will 

be clamped on the drawer. When the drawer reaches its end position inside the 

Incubator Shaker, the labware will sink on the heat plate for best thermal contact. You 

can hear a short shaking noise to ensure, that the labware is sinking on the heat plate 

but only when the labware is detected correctly from the device.  

4.7.1. 

Incubator Mp and Incubator shaker Mp

The Incubator MP and Incubator Shaker MP are built for the use with microtiterplates. 

The Labware detection is optimized for Greiner Plates. If different microtiterplates are 

used, the user has to assure, that the Labware detection works well with these plates 

too. 

4.7.2. 

Incubator DWP and Incubator Shaker DWP

The Incubator DWP and Incubator Shaker DWP are built for the use with deep well 

plates. The Labware detection is optimized for Nunc Plates. If different deep well plates 

are used, the user has to assure, that the Labware detection works well with these 

plates too. 

4.7.3. 

Criteria for labware determination

-  Light master plates are better than dark ones
-  Master plates with matt surface are better than shiny ones
-  Attaching an adapted tape on the left side of the master plate improves the success 

rate of the Labware detection.  If you want to write your own software control or 

integrate it in an existing workstation conrol keep you should keep in mind the 

following fact. With the command ASE1 the incubator shaker is unable to shake 

without detected Labware. The command ASEND has been added to enable shaking 

without detected Labware. We do not recommend using it! The command ASE1 is 

preferred. It ensures that the Labware is correctly placed on the shaker and performs 

within the specifications.

Содержание DWP

Страница 1: ...Doc ID 900434 001 March 2018 INHECO Single Position Incubators Incubator MP DWP Part No 7300003 7300006 Incubator Shaker MP Shaker DWP Part No 7300013 7300009 User s Manual...

Страница 2: ...nual This manual and the information herein have been assembled with due diligence INHECO GmbH does not assume liability for any misprints or cases of damage resulting from misprints in this manual If...

Страница 3: ...tails General 18 Adress Switch 19 Connecting devices 19 Mechanical Integration 21 Labware Use and Detection 22 Communication with the Incubator 23 Integration in workstation 23 COM port identification...

Страница 4: ...s meet the acknowledged rules of technology and comply with today s standards Manual instructions must be followed in order to limit the safety risk during operation of the unit Security related warni...

Страница 5: ...roller Microprocessor with on chip peripheral CRC Cyclic Redundany Check FCS Firmware Command Set Heater Same as Temperature Controller IVD D In vitro diagnostic directive MP Micro Plate PCR Polymeras...

Страница 6: ...t have been used in Biosafety Level 3 and 4 environments or have been exposed to radioactive or radiation materials Cleaning and Decontamination page 37 Devices exposed to Biosafety level 3 and 4 Envi...

Страница 7: ...lling the devices enabling the use of the Incubator products in FDA IVD compliant environments and applications Software and verification tools are available to form a complete thermal solution for al...

Страница 8: ...and products and email them to TechHotline inheco com immediately Transportation damages must be reported to INHECO within 7 days of delivery The following components should be included in each shipm...

Страница 9: ...ents Figure 2 introduces the terminology of the Incubator s functional elements Fig 2 Functional elements 1 Fan 4 USB Connnector 7 15 pole Sub D Connector male 2 Loading drawer 5 Address Switch 8 15 p...

Страница 10: ...section of www inheco com Manual INHECO Incubator Shaker control Incubator MP Incubator DWP 1000 INHECO GmbH 82152 Martinsried Germany 04 PN SN Input 24V 7300003 Incubator MP 2016 09 Rev 30W Fuse F3...

Страница 11: ...g Shaker frequency range N A 6 6 Hz to 30 HZ 400 rpm to 1800 rpm Maximum shaking amplitude difference between the maxima N A 3 mm Shaking direction N A Arbitrary programmable e g orgital linear Dimens...

Страница 12: ...5 kg Open drawer resistance for a force in Z axis without plastic deformation 10 N Enviromental Conditions Maximum operation height Tolarable relative humidity Operation 10 80 relative non condensing...

Страница 13: ...East West Shaker Load 100g Shaker Speed Hz Shaker amplitude mm Single device Maximum stack 10 X 1 0 to 2 2 or Y 1 0 to 2 2 X 1 0 to 1 9 or Y 1 0 to 1 9 15 X 1 0 to 2 7 or Y 1 0 to 2 7 X 1 0 to 2 4 or...

Страница 14: ...ture according to a Greiner 96 Flat Bottom Plate Postitions A1 A12 E7 H12 Target Temp Validation Tool Uniformity C 37 C 98 6 F 1 8 Evaporation in unsealed Micro Plates can be reduced by using lid or c...

Страница 15: ...nments with risk of explosion The unit is for indoor use only The unit must not be used if the Incubator the power adapter housing or the power cable show visible signs of damage Make sure that the el...

Страница 16: ...SD electrostatic discharge sensitive devices Electrostatic charges as high as 4000V accumulate on the human body can discharge without detection Although the Incubator devices feature proprietary ESD...

Страница 17: ...the Incubator 4 2 1 Disassembling the transportation screws only for Shaker devices After unpacking the Incubator Shaker device the two red screws transportation lock at the bottom side have to be uns...

Страница 18: ...ttom of the device Turn the Incubator Shaker slowly to the left Note Cover side in direction of assembler like in figure 10 unscrew the screw just by hand Fig 9 Screws screwable without screwdriver Af...

Страница 19: ...not optimized for a secure real time data transfer all communication is secured by a cyclic checksum CRC In case the communication between the computer and the Incubator Unit often fails due to timeo...

Страница 20: ...wers INHECO s compact single position Incubators are high performance plug play devices for the use on automated robotic platforms or as standalone units They can either be used as single devices or s...

Страница 21: ...figure 12 on page 21 Slave unit needs no USB cable or power cord Two connected devices 4 5 3 How to connect two devices slave to slave plug in the 15 pole SUB D Connectors The 15 pole SUB D Connector...

Страница 22: ...te plate for a firm and even stand please contact your workstation provider to find out if they offer an appropriate adapter plate Use the position pins on the bottom side for exact positioning of the...

Страница 23: ...correctly from the device 4 7 1 Incubator MP and Incubator Shaker MP The Incubator MP and Incubator Shaker MP are built for the use with microtiterplates The Labware detection is optimized for Greiner...

Страница 24: ...optimized for a secure real time data transfer all communication is secured by a cyclic checksum CRC In case the communication between the computer and the Incubator Unit often fails due to timeouts i...

Страница 25: ...hould be found in the list of assigned ports The COM port is shown behind the the USB Serial Port in brackets Please continue with start the workstation software Additional dialogues could appear depe...

Страница 26: ...o a robotic systems Please contact the supplier of your robotic system to discuss the integration routine Atypical cycle for a process could contain the following steps Scan COM Ports mandatory Find d...

Страница 27: ...r different setup conditions Some devices may be already connected to the USB green LED is on some may not green LED is blinking at 1 Hz The blinking of the LED stops only after the first message with...

Страница 28: ...et with your workstation software Manual Workstation Software NOTE Only for Incubator MP device The default Heat up Off set for 37 C is 4 C for a Greiner 96 Flat Bottom Plate If your workstation softw...

Страница 29: ...ide gas The surface decontamination can include a wipe down of the housing with a moistened cloth Ethanol 70 can be used where effective against target organisms The use of Bleach 5 12 or Microside SQ...

Страница 30: ...O for support INHECO product number of the device shown on device label INHECO product name of the device shown on device label INHECO serial number of the device shown on device label or via software...

Страница 31: ...environments are not accepted by INHECO for return Ask techhotline inheco com or visit www inheco com service returns rma html for the return procedure before you return a device to INHECO Do not retu...

Страница 32: ...on the lock correctly With this movement the front rear shaker is moved to the back and only then the lock can be placed correctly 2 Turn the unit about 45 60 to the left to position the lock correctl...

Страница 33: ...ant check of all data is done The INHECO Measurement Plate IMP can be used for verifation of shaking amplitude and temperature 8 3 Automatically self test AQS The automatically self testAQS which is o...

Страница 34: ...contact the support of your workstation provider 9 1 Drawer failure 9 1 1 error 10 check whether the labware matches the labware criteria specified in chapter xxx check whether the opening for the ele...

Страница 35: ...re clean check whether the seal ring of drawer is in correct position Fig 29 contact area on drawer lid and correct seal ring position Fig 30 contact pins reboot the incubator 9 1 3 drawer is not clos...

Страница 36: ...tails INHECO Part Number Power Supply built for 1 and more incubators AC Input 100 240 V 2 5A 50 60 Hz and DC Output 24 V 7 5 A 180W 2400121 10 2 Power Supply Cord 10 2 1 Order Details INHECO Part Num...

Страница 37: ...INHECO 36 Doc ID 900434 001 March 2018 11 Appendix...

Отзывы: