INHECO
12
Doc ID: 900434-001
March 2018
2.6.
shaker functionality Incubator shaker Mp / DWp
Parameter
Value
Amplitude range
0.0 – 3.0 mm*
Speed range
6.6 – 30.0 Hz*
Speed precision
±2 Hz
Load
40 – 250g* (weight of plate and liquid and adapter)
Supervision
speed and amplitude (closed loop controller)
Shaker movement
North – South (linear)
East – West (linear)
Round (orbital)
“Eight” (this is possible but not recommended because it might damage the
shaker)
Shaker Movement
Amplitude accuracy**
±20% (of the target amplitude)
Phase shift accuracy**
±11°
** validated for the Heating Plate movement / average value over 10 cycles
Single Axis (North – South and East – West)
Shaker Load: 100g
Shaker
Speed [Hz]
Shaker amplitude [mm]
Single device
Maximum stack
10
X = 1.0 to 2.2
or
Y = 1.0 to 2.2
X = 1,0 to 1.9
or
Y = 1.0 to 1.9
15
X = 1.0 to 2.7
or
Y = 1.0 to 2.7
X = 1.0 to 2.4
or
Y = 1,0 to 2.4
20
X = 1,0 to 1.2
or
Y = 1.0 to 1.2
X = 1.0 to 1.2
or
Y= 1.0 to 1.2
round (phase shift = 90°)
shaker Load: 100g
Shaker
Speed [Hz]
Shaker amplitude [mm]
Single device
Maximum stack
10
X = 1.0 to 2.2
Y = 1.0 to 2.2
X = 1,0 to 1.9
Y = 1.0 to 1.9
15
X = 1.0 to 2.7
Y = 1.0 to 2.7
X = 1.0 to 2.4
Y = 1.0 to 2.4
20
X = 1.0 to 1.2
Y = 1.0 to 1.2
X = 1.0 to 1.2
Y = 1.0 to 1.2
„eight“ (phase shift = 0°)
shaker Load: 100g
Shaker
Speed [Hz]
Shaker amplitude [Hz]
Single device
Triple Stack
X = 12 /
Y = 24
X = 1.0 to 2.4
Y = 1.0 to 2.4
X = 1.0 to 2.2
Y = 1.0 to 2.2
X = 15 /
Y = 30
X = 1.0
Y = 1.0
X = 1.0
Y = 1.0
* For detailed shaker performance, see chapter below.
X
Y
front
shaking table