12 - 1
Transpector XPR 3+ Operating Manual
Chapter 12
Specifications
12.1 Introduction
The following sections detail the specifications for Transpector XPR 3+ Gas
Analysis System. As a result of INFICON continuing product improvement and
quality assurance programs, these specifications are subject to change without
notice or obligation.
NOTE:
All specifications are measured after a 30 minute warm-up period at
constant STP unless specified.
12.2 Mass Range
12.3 Detector Type
12.4 Resolution
12.5 Temperature Coefficient
Transpector XPR 3+
Transpector XPR 3+ Low Pressure EM
1 to 100 AMU
Transpector XPR 3+
Transpector XPR 3+ Low Pressure EM
Microchannel Plate Electron
Multiplier/ Faraday Cup
Continuous Dynode Electron Multiplier /
Faraday Cup
Peak Width (AMU) at 10% of peak height
Transpector XPR 3+
Transpector XPR 3+ Low Pressure EM
1.0 AMU
During an eight hour period, after thirty minutes of warm up
Transpector XPR 3+
Transpector XPR 3+ LP
For FC Only—Less than 1% of peak height per °C