2 - 12
Transpector XPR 3+ Operating Manual
2.4.2.1 Installing Multiple Transpector XPR 3+ Directly to a Host Computer
If multiple Transpector XPR 3+ sensors are to be connected to a single host
computer and not to an existing local area network, a private local network must be
created. Transpector XPR 3+ will have to be installed on either a router or Ethernet
switch. The router or switch is then connected to the host computer through the
LAN port of the router/switch.
2.4.2.2 Installing Multiple Transpector XPR 3+ on an Existing Local Network
If multiple Transpector XPR 3+ sensors are to be connected to an existing local
network, use an Ethernet switch instead of a router.
NOTE:
Routers can cause conflicts with local networks because the router will
attempt to set IP addresses for all network connected devices.
Since Transpector XPR 3+ sensors will be network connected devices, each
sensor must have an IP address assigned to it by a network administrator. After
changing each IP address manually, connect all of the sensors to the Ethernet
switch and connect the switch to the local network.