1 - 21
Transpector XPR 3+ Operating Manual
CAUTION
The silver-plated bolts used for mounting the sensor to
the vacuum system must be oriented such that the bolt
heads are on the same side of the sensor as the
electronics box. Otherwise, there may be interference
between the black INFICON Transpector mounting nut
and sensor mounting hardware.
CAUTION
Maximum bakeout temperature for sensors is shown in
CAUTION
Transpector XPR 3+ electronics module must be
removed prior to bakeout at temperatures greater than
150°C (FC).
Do not turn on the Electron Multiplier (EM) at sensor
temperatures above 150°C. Turning on the EM at elevated
temperature could result in permanent damage to the
detector.
WARNING
During or immediately after bakeout, the heating jacket
and metal surfaces in the vicinity of the heating jacket
may be extremely hot. These surfaces may exceed 100°C
at the maximum ambient operating temperature (50°C),
which will cause burns if touched directly without using
the proper personal protection equipment.
Table 1-1 Sensor maximum bakeout temperature
Sensor
Maximum
Operating
Temperature
Maximum Bakeout
Temperature
Electronics Removed
Transpector XPR 3+
150°C
200°C