7 - 7
Transpector XPR 3+ Operating Manual
Figure 7-4 Filament in place on sensor
4
While holding the short end of the hex wrench, alternately tighten the screws
until the hex wrench begins to flex. Alternately use a torque-limiting screwdriver
(IPN 02-389-P1 or equivalent), torque the screws to 10-12 oz In. (0.0384 Nm
to 0.0461 Nm).
CAUTION
Do not touch the filament wires.
7.5.4 How to Determine the Condition of the Ion Source
1
Remove the Transpector XPR 3+ sensor from the vacuum system.
2
Completely loosen the two 1-72 x 0.31 in. long gold plated cap head screws
that hold the filament assembly to the ion source plate. Remove the filament
assembly and screws. See
Face of repeller parallel to face of anode