8 - 3
Transpector XPR 3+ Operating Manual
Temperature error
Transpector XPR 3+ internal
ambient temp >70°C
Verify unit is installed
properly, ambient temp
<50°C.
Verify that there are no heat
sources in local proximity.
Verify that internal fan is
running.
Electronics failure
Return to INFICON for repair.
Overpressure
Total pressure plate current
exceeded trip threshold
Reduce pressure.
Total pressure plate is
contaminated
Return to INFICON for repair.
Electrometer Error
Excessive current (>2E-6
Amps)
Reduce pressure.
Reduce EM voltage.
Electronics failure
Return to INFICON for repair.
Mass Filter Error
Bad tune file
Retune RGA.
Contaminated quadrupole
Return to INFICON for repair.
Ion Source Error
Incorrect Ion source settings
Load original configuration
file.
Ion source contaminated
Return to INFICON for repair.
Detector Error
EM voltage set too high
Reduce EM voltage.
Detector not working
correctly
Replace EM detector.
DSP
Communications
Error
Incorrect DSP response
detected
Reset electronics box.
Return electronics box to
INFICON for repair.
DEC
Communications
Error
Incorrect DEC response
detected
Reset electronics box.
Return electronics box to
INFICON for repair.
Table 8-1 Symptom—Cause—Remedy chart (continued)
SYMPTOM
CAUSE
REMEDY