4.6 Initial Power-On Verification
A preliminary functional check of IMM-200 can be made before installation. It is not
necessary to have sensors or communication cables connected to do this.
The green IMM-200 Status light is used to identify if the unit is operating correctly, or if
there is a failure condition.
Connect 24 V (dc) to the power connector using either one of the power supplies listed
in Parts and Options Overview or a user supplied power supply with a power rating
indicated by Power, in Specifications.
If a user supplied power supply is used, INFICON recommends the power connection
to IMM-200 use a Kycon connector, part number KPPX-4P. Pins 3 and 4 carry +24 V
(dc), pins 1 and 2 are ground. Whenever a user supplied power supply is used, the
connector shield on the 24 V power cable must be connected to the mains protective
earth ground terminal through the user supplied power supply.
If a good crystal is connected, the Status LED illuminates and remains steadily on.
This status indicates proper operation of IMM-200. If a good crystal is not connected,
the Status LED illuminates 1 s on,1 s off, to indicate a crystal fail. IMM-200 is in Idle if
the Status LED flash pattern is 1 s on, 4 s off.
If the Status LED has the following flash patterns, please contact INFICON as an error
exists. To contact INFICON service, refer to How to Contact Us [
LED State
Description
0.5 s on, 0.2 s off, 0.5 s on, 1 s off
Invalid firmware or POST fail
If the Status LED is off, confirm that the proper power is connected to IMM-200. If the
proper power is connected but the LED is still off, contact INFICON service.