7 Ethernet Communications
7.1 Introduction
IMM-200 uses a Lantronix
®
XPort Ethernet communications interface for sending
parameter values, instrument commands, and status, warning and error messages
between IMM-200 and a host computer. INFICON P3Plus communications protocol
packets with computed CRC are transferred between the IMM-200 and the Lantronix
adapter (8, N, 1, 921600 bps).
7.2 Setting the IP Address
IMM-200 Ethernet settings can be configured using the Lantronix DeviceInstaller
Windows application downloadable from: https://www.lantronix.com/products/
deviceinstaller/
There are a few ways to set the IP address. All of them are described in the User
Guide for the XPort device. https://cdn.lantronix.com/wp-content/uploads/pdf/
DeviceInstaller_UG.pdf
If the IP address is already known, type it into the browser. A prompt appears with a
login dialog. IMM-200 is shipped with both the username and password unset/empty,
so leave them blank and click OK to login. The IP address for IMM-200 is under
Network. Make sure that the IP address for the Remote Host (under Channel 1 |
Connection) is set to the same subnet as the new IP. It is not used, but the IP must be
compatible.
If the IP address is unknown, the easiest way to access the device is to install the
Lantronix DeviceInstaller Windows application. It will find all Lantronix devices on the
network. From the factory the default configuration of the Ethernet module is set for an
IP address of 192.168.1.101 and a subnet mask of 255.255.255.0. The DeviceInstaller
may find many Lantronix devices on the network. IMM-200 uses the XPort-05. Once
the XPort device to be modified is selected, use the Web Configuration from within the
app or just use the browser (with the known IP address).
7.3 Protocol
IMM-200 uses the P3Plus (P3P) communication protocol with the following notes:
• ADR byte not evaluated (set to 0)
• ADR byte included in CRC calculation
• DEV ID ignored by slave on Requests to slave.