INFICON
Troubleshooting and Error Messages | 9
074-733-P1A IMM-200 Operating Manual
47 / 67
SYMPTOM
CAUSE
REMEDY
c. Check the sensor
magnetic field strength.
The maximum field at the
center of the aperture
should be 700 gauss or
greater.
8. There is a rotary sensor
crystal switch problem.
(The sensor does not
advance or is not centered
in aperture.)
a. There is no relay or an
incorrect relay output
programmed (for
instruments having
outputs).
b. There is a loss of
pneumatic supply or the
pressure is insufficient for
operation.
c. Operation has been
impaired as a result of
material accumulation on
the cover.
d. There is an improper
alignment.
e. A 0.057 mm (0.00225
in.) diameter orifice is not
installed on the supply side
of solenoid valve assembly
(CrystalSix or Crystal12
sensors).
a. Program a relay.
b. Ensure the air supply is
regulated at 80-90 psi.
c. Clean material
accumulation as needed.
d. Refer to the sensor
operating manual for the
alignment procedure.
e. Install the orifice as
shown in the CrystalSix or
Crystal 12 Operating
manual.
9.3 Troubleshooting Computer Communication
Symptom
Cause
Remedy
1. Communication cannot
be established between
the computer and
IMM-200.
a. There is an improper
communication cable
connection.
b. There is a defective
communication cable.
c. The incorrect IP address
or submask being used.
a. Ensure that the cables
are connected to the
correct ports on IMM-200
and the computer.
b. Verify that the cable is
plugged in and is not
damaged. Replace the
communication cable.