25 / 67
LED State
Description
boot loader flash
1
(0.4 second on,
0.4 second off)
Normal, a valid firmware version is present.
boot loader flash
2
(0.2 second on,
0.2second off)
Old firmware is being overwritten, new firmware has
been downloaded and passed the integrity check.
boot loader flash
3
(0.5 second on,
.2 second off,
.5 second on,
1 second off)
The firmware is invalid.
off
The unit is in diagnostics mode, or the unit has
encountered an error, or no power is applied to the
device.
flash
(1 second on
4 seconds off)
The unit is in idle mode.
blinking
(1 second on
1 second off)
The unit is in Measure mode and a crystal failure has
been detected.
on
The unit is in Measure mode
or the unit is running normally in boot.