13 - 13
IP
N 07
4-
39
7-
P1
G
HAPSITE Smart Operating Manual
13.2.5 Making the Measurement
To make a measurement using the HSS, a method is required. To quantify the
analytes detected, a calibrated method is required.
Once the calibrated method is completed, samples may be analyzed and
compared to the calibration for quantitative analysis. Sample analysis must be
performed using the same procedure used to perform the calibration runs.
Chapter 12, Target Compound Methods
for additional information on developing methods and calibration for quantitative
analysis.
13.3 Maintenance of HSS
Other than replacement of consumable items, most maintenance will involve
routine purging of the system as needed when analyzing high concentration
samples, needle replacement, or cleaning of the sample wells. Refer to
15.10, Replacing the HSS Needle, on page 15-51
13.3.1 Clean Out after a High Concentration Sample
When high concentration samples are analyzed using the HSS, the system can
retain analytes which will be detected if the next sample has low levels of those
compounds. This is typically referred to as a carry-over contamination. Carry-over
contamination can be reduced to approximately 0.1% by using
Purge
. In the case
of severe carryover, use the
Flush
function.
13.3.1.1 Purge Procedure
To
purge
the HSS, run a blank with the HSS needle in a clean, empty purge vial.
In some cases, the blank may need to be run more than once.
HINT:
A blank is the same method run without introducing a sample.
13.3.2 Cleaning Up a High Background
High background is usually encountered because some compounds are retained
in the HSS or the HAPSITE. This could be in the HSS valves, transfer line, or within
the HAPSITE.
To test whether the background is in the HSS or the HAPSITE, disconnect the
Transfer Line, connect the air sampling probe to the HAPSITE, and analyze a clean
air sample. If the contamination is no longer present, the contamination is in the
HSS or Transfer Line.
Содержание HAPSITE Smart
Страница 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Страница 2: ......
Страница 12: ......
Страница 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 107: ...3 19 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 24 Selecting Survey Method using Recent Methods Menu ...
Страница 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Страница 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 359: ...11 21 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 19 Method Editor Tune Parameter Page ...
Страница 360: ...11 22 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 20 Method Editor Tune Parameter Page Show Details ...
Страница 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Страница 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 476: ...15 18 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 15 13 Vent HAPSITE from Smart IQ Software ...
Страница 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 548: ......