HAPSITE Smart Operating Manual
Clear All Empty Compound Entries
. . . . . . . . Deletes all entries that do not
have a compound name associated with
them.
NOTE:
The down arrow next to the compound name can be used to select a
different name if more than one possible match has been determined
by the AMDIS search.
9a
For this file, all of the blank entries are to be deleted. Click the right mouse
button in the
Compound
field and select
Clear All Empty Compound
Entries
. Refer to
HINT:
If a compound is unidentified (blank), but is known to the analyst, name
the peak by highlighting the blank compound name field and entering
the name.
9b
Next, the Internal Standards must be selected. When using Internal Standards,
best practice is to use a quant Ion from the Internal Standard that is close in
mass to the quant Ion of the compound to be quantitated. The software always
selects the largest mass fragment in the spectrum as the quant Ion. To change
the quant ion, highlight the field and type in the new number. Change the quant
Ion for TRIS from
213
to
75
. The software will warn that a re-calibration is
required. This is true if a quant ion is changed after calibration. To signify which
quant mass is being used, change the name of
TRIS HAPSITE IS #1
to
TRIS_75
NOTE:
If the name of the analyte begins with TRIS or BPFB the software will
recognize the analyte as a HAPSITE IS and automatically enter the
concentration from the IS canister into the method for calibration and
quantitation.
9c
More than one quant ion can be used from a single Internal Standard peak.
Highlight the second Internal Standard and click the right mouse button, select
Duplicate Row
, then change the name of the Internal Standard peaks to
BPFB_79
and
BPFB_117
. Next the
Quant Ion
should be changed to
79
and
117
. See
.
Figure 12-16 Finished Analytes Chart
Содержание HAPSITE Smart
Страница 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Страница 2: ......
Страница 12: ......
Страница 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 107: ...3 19 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 24 Selecting Survey Method using Recent Methods Menu ...
Страница 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Страница 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 359: ...11 21 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 19 Method Editor Tune Parameter Page ...
Страница 360: ...11 22 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 20 Method Editor Tune Parameter Page Show Details ...
Страница 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Страница 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 476: ...15 18 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 15 13 Vent HAPSITE from Smart IQ Software ...
Страница 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 548: ......