3 - 30
IP
N 07
4-
39
7-
P1
G
HAPSITE Smart Operating Manual
Figure 3-40 Front Panel Run Prompt
5
Press the
RUN
button on the Front Panel or on the Probe. This run will take
approximately three minutes to complete. The progress of the run can be seen
on the Front Panel. See
NOTE:
The
MAX
_
TIC
number on the top left of the LCD must be less than
500,000 for the concentrator tube to be considered clean. If the
MAX
TIC
number is greater than 500,000, repeat the 3 min Carbon Conc
Cleanout method until the
MAX_TIC
number becomes less than
500,000. Please note, in
the MAX_TIC is 1,551,539
therefore, the concentrator is NOT clean. The 3 min Carbon Conc
Cleanout would need to be run again in this example.
Figure 3-41 Front Panel 3 min Carbon Conc Cleanout Run
6
When the 3 min Carbon Conc Cleanout method is completed and the
MAX_TIC
count is less than 500,000, press
ESC
to exit the method. The system is now
clean enough to run the
Air 15 min Carbon Conc
method. See
.
Содержание HAPSITE Smart
Страница 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Страница 2: ......
Страница 12: ......
Страница 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 107: ...3 19 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 24 Selecting Survey Method using Recent Methods Menu ...
Страница 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Страница 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 359: ...11 21 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 19 Method Editor Tune Parameter Page ...
Страница 360: ...11 22 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 20 Method Editor Tune Parameter Page Show Details ...
Страница 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Страница 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 476: ...15 18 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 15 13 Vent HAPSITE from Smart IQ Software ...
Страница 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 548: ......