HAPSITE Smart Operating Manual
3.7.1 Quick Reference SOP —
GC/MS Mode with HSS and Sample Loop in Portable Mode
1
From the Front Panel highlight
Run Method
, then press
SEL
.
2
Highlight
GC/MS
, then press
SEL
.
3
Select the
Headspace
folder from the GC/MS menu.
4
Select
HSS 15 Min Loop
method
5
Place the headspace needle in the 40mL vial that contains the sample.
6
When prompted, press the
Run
button on the front panel of the HAPSITE.
7
When prompted, place the headspace needle in the purge vial and press
SEL
.
8
Review results.
3.8 GC/MS Mode with Headspace Sampling System and
Concentrator in Portable Mode
This method is used for analyzing samples with concentration levels in the low part
per trillion range. This method is used when the loop method fails to detect a
chemical. Refer to
for additional information on GC/MS
methods.
CAUTION
The concentrator feature is designed to enhance
HAPSITE sensitivity when analyzing low concentration
samples. The accumulated sample VOCs are transferred
from the concentrator adsorbent bed to the HAPSITE as
a concentrated sample. Care should be taken to avoid
overloading the HAPSITE and causing excessive
carryover of VOCs in subsequent runs.
1
Refer to
Chapter 13, Headspace Sampling System
for more information on the
Headspace System. Refer to
2.6, HAPSITE Components and Assemblies
for
assembly instructions.
2
Follow the instructions in
Section 2.9.3, Installing the Tri-Bed Concentrator, on
Section 2.9.4, Installing the Tenax or Carbopack Concentrator, on
to install the concentrator tube.
3
To cleanout out the concentrator tube, follow the instructions in
Quick Reference SOP — Tri-Bed Concentrator Cleanout, on page 3-32
; or
Section 3.6.1, Quick Reference SOP — Tenax Cleanout, on page
3-35
;depending on the type of concentrator installed.
Содержание HAPSITE Smart
Страница 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Страница 2: ......
Страница 12: ......
Страница 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 107: ...3 19 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 24 Selecting Survey Method using Recent Methods Menu ...
Страница 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Страница 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 359: ...11 21 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 19 Method Editor Tune Parameter Page ...
Страница 360: ...11 22 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 20 Method Editor Tune Parameter Page Show Details ...
Страница 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Страница 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 476: ...15 18 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 15 13 Vent HAPSITE from Smart IQ Software ...
Страница 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 548: ......