HAPSITE Smart Operating Manual
13.3.6 Replacing The HSS Nitrogen Canister
Replacement of the Nitrogen Canister is on an as needed basis. When the
pressure is below approximately 140 kPa, the canister will need to be replaced with
a new Nitrogen Canister. Replacement should be performed between sample runs,
and in a dry and non-hazardous environment, since the front door must be opened.
To change the Nitrogen Carrier Gas Canister:
1
Open the front door of the HSS and locate the Nitrogen Canister.
2
Press the black release tab on the left side of the Nitrogen Canister and the
canister will partially pop out.
3
Remove the canister from the slot.
4
Remove the plastic protective cap from a new Nitrogen Canister and insert into
the canister slot.
5
Press and hold the black release tab while pushing the Canister until the
canister stops, then release the tab. The canister should be snug and locked in
place.
WARNING
Do not re-fill the canisters after use. These canisters are
disposable and not designed for re-filling. Canisters may
fail upon refilling, causing bodily injury.
13.4 Shipping the HSS
The HSS can be readily shipped to a field location. The cardboard box in which the
HSS was received, together with the cut-foam end-caps can be reused. A
reinforced plastic shipping case (IPN 931-406-P1) is also available.
Before shipping the instrument, place an empty vial in one of the wells to insert the
needle in, for protection. Remove the battery and the canister. No other steps are
required to prepare the HSS for shipment.
Содержание HAPSITE Smart
Страница 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Страница 2: ......
Страница 12: ......
Страница 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 107: ...3 19 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 24 Selecting Survey Method using Recent Methods Menu ...
Страница 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Страница 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 359: ...11 21 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 19 Method Editor Tune Parameter Page ...
Страница 360: ...11 22 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 20 Method Editor Tune Parameter Page Show Details ...
Страница 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Страница 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 476: ...15 18 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 15 13 Vent HAPSITE from Smart IQ Software ...
Страница 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 548: ......