HAPSITE Smart Operating Manual
Figure 12-19 Library Ready for Calibration
11a
Prior to calibration, the curve fit should be selected. To change the curve fit from
the default
Linear, Forced through Origin
, select the
View/Edit
button under
the
Method
controls. Select the curve fit and save the method.
11b
To calibrate the library, select the standard data files, select
Calibrate Library
and press the
Start
button under
Peak Search
.
11c
Review the curves for each analyte. To review each analyte, select the
Calibration
tab and use the mouse to select each compound. See
.
Figure 12-20 Calibration Curve
11d
Additional calibration points can be added to the curve at any time by using the
Browse
button under
Data Files
. First add the data file. Select the data file
added and make sure that the previous data file is no longer selected by
un-checking the select field. Select
Start
under
Peak Search
. Review the
curves as with the previous data file. Each curve should have all of the
calibration points.
Содержание HAPSITE Smart
Страница 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Страница 2: ......
Страница 12: ......
Страница 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 107: ...3 19 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 24 Selecting Survey Method using Recent Methods Menu ...
Страница 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Страница 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 359: ...11 21 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 19 Method Editor Tune Parameter Page ...
Страница 360: ...11 22 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 20 Method Editor Tune Parameter Page Show Details ...
Страница 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Страница 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 476: ...15 18 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 15 13 Vent HAPSITE from Smart IQ Software ...
Страница 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 548: ......