HAPSITE Smart Operating Manual
CAUTION
The concentrator feature is designed to enhance
HAPSITE sensitivity when analyzing low concentration
samples. The accumulated sample VOCs are transferred
from the concentrator adsorbent bed to the HAPSITE as
a concentrated sample. Care should be taken to avoid
overloading the HAPSITE and causing excessive
carryover of VOCs in subsequent runs.
3.6.1 Quick Reference SOP — Tenax Cleanout
1
From the main menu on the Front Panel highlight
Run Method
and press
SEL
.
2
Highlight
GCMS
and press
SEL
.
3
Highlight
Concentrator
and press
SEL
.
4
Highlight
3 min Tenax Conc Cleanout
and press
SEL
.
5
Wait for heaters to reach set temperatures.
6
When prompted press
RUN
from the Front Panel or Probe.
NOTE:
If the
TIC_Max
count is less than 500,000 at the end of this run, proceed
to the
Air 15 min Tenax Conc
method. If the
TIC_Max
count is greater
than 500,000, rerun the
3 min Tenax Conc Cleanout
method. Repeat this
until the
TIC_Max
is less than 500,000.
3.6.2 Quick Reference SOP — Tenax Concentrator Method
1
The 3 min Tenax Conc Cleanout method must be run prior this method. Refer
to
Section 3.6.1, Quick Reference SOP — Tenax Cleanout, on page 3-35
.
2
From the main menu on the Front Panel highlight
Run Method
and press
SEL
.
3
Highlight
GCMS
and press
SEL
.
4
Highlight
Concentrator
and press
SEL
.
5
Highlight
Air 15 Min Tenax Conc
and press
SEL
.
6
Wait for heaters to gain control of set temperatures.
7
When prompted press
RUN
from the Front Panel or Probe.
8
Hold the probe over the sample for both the one minute
LinePurge
and the one
minute
Concfill
event.
9
Wait for run to complete to view results.
Содержание HAPSITE Smart
Страница 1: ...O P E R A T I N G M A N U A L HAPSITE Smart Chemical Identification System IPN 074 397 P1G ...
Страница 2: ......
Страница 12: ......
Страница 28: ...TOC 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 88: ...2 50 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 107: ...3 19 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 24 Selecting Survey Method using Recent Methods Menu ...
Страница 139: ...3 51 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 3 75 Shutting Down the HAPSITE ...
Страница 160: ...4 18 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 286: ...8 42 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 324: ...9 38 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 338: ...10 14 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 359: ...11 21 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 19 Method Editor Tune Parameter Page ...
Страница 360: ...11 22 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 20 Method Editor Tune Parameter Page Show Details ...
Страница 390: ...11 52 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 11 55 Saving Sequenced Method ...
Страница 412: ...12 22 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 432: ...13 20 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 458: ...14 26 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 476: ...15 18 IPN 074 397 P1G HAPSITE Smart Operating Manual Figure 15 13 Vent HAPSITE from Smart IQ Software ...
Страница 528: ...16 12 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 530: ...17 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 538: ...A 2 IPN 074 397 P1G HAPSITE Smart Operating Manual This page is intentionally blank ...
Страница 548: ......