Cheetah Pregius Camera with 10 GigE Vision Interface – User Manual
October 8, 2020
Page
120
of
170
Rev 1.0
Standard Trigger Mode
In
Standard
Trigger mode
, the camera first performs the exposure and then reads out the image.
An external timing pulse controls the start of the exposure if exposure control mode is set to
Timed
.
Trigger
Exposure time
Exposure N
Exposure N+1
N
N+1
Readout N
Readout N+1
Readout time
Ignored
Ignored
Trigger period
GenICam controls
TriggerMode:
On
TriggerOverlap:
Off
TriggerSource:
Line 1
(or Line2, Software, Pulse Generator)
TriggerActivation:
Rising Edge
(or Falling Edge)
TriggerFilterTime, TriggerDelay, TriggerDebounceTime: –
set if applicable
.
Exposure Mode:
Timed
For manual exposure control:
Exposure Time:
User-specified
(Min.= 36
µ
s; Max = Readout time)
ExposureAuto:
Off
For automatic exposure control (AEC):
ExposureAuto:
Continuous
(or Once)
Parameters of the external trigger pulse
Trigger width:
≥ 10
µ
s
Trigger period (min) = Exposure time (max) + Readout time
Start of
Exposure
Standard trigger mode, Exposure control is Timed
Trigger active
edge – Rising
Trigger
width
Exposure time