G P I B C O M M A N D R E F E R E N C E
GPIB Command Reference
54
FPM-8200
C H A P T E R
4
Action
Requests an ASCII string from FPM−8200 message memory.
Response
16−character string, followed by NULL terminator.
Note
As necessary, the string is filled with spaces to 16−character length. If no message string was
stored, the response is a string of 16 spaces.
Example
Message?
Response:
This response means the previously stored message was Test 3.
Action
Selects the units to be used in reporting measurements:
DB
Decibels (dB) relative to the reference value.
DBM
Logarithmic (dBm) units.
W
Linear (watts) units.
Notes
dBm is the log of the ratio of output power relative to one milliwatt. The calculation is:
MODE defines the units of value that will be used for responses to a POW? request.
This command has the same effect as selecting the display mode on the front panel.
Example
Mode:dB˙Measurements are reported in dB (decibels) relative to the reference value.
MESsage?
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
MODE:
DB
DBM
W
C
OMMON
D
EVICE
D
EPENDENT
F
RONT
P
ANEL
Test 3 <NULL>
16 characters total
Fill spaces
output power in milliwatts
1 milliwatt
dBm = 10 log
Содержание FPM-8200
Страница 2: ......
Страница 8: ...L I S T O F FIG U R E S vi FPM 8200...
Страница 10: ...L I S T O F TA B L E S viii FPM 8200...
Страница 16: ...WA R R A N T Y xiv FPM 8200...
Страница 36: ...FR O NT PA N E L O P E R A T I O N Operating Guidelines 20 FPM 8200 C H A P T E R 2...
Страница 90: ...TR O U B L E S H OO T I N G A N D M A I N T E N A N C E Error Messages 74 FPM 8200 C H A P T E R 5...
Страница 96: ...80 FPM 8200 A P P E N D I X A...